Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
Measuring Reaction Rates03:09

Measuring Reaction Rates

Polarimetry finds application in chemical kinetics to measure the concentration and reaction kinetics of optically active substances during a chemical reaction. Optically active substances have the capability of rotating the plane of polarization of linearly polarized light passing through them—a feature called optical rotation. Optical activity is attributed to the molecular structure of substances. Normal monochromatic light is unpolarized and possesses oscillations of the electrical field in...
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Gas Chromatography: Types of Detectors-I01:21

Gas Chromatography: Types of Detectors-I

There are different types of detectors used in gas chromatography, each with its own specific properties that make it suitable for detecting certain types of analytes. The most commonly used detectors in GC are thermal conductivity detector (TCD), flame ionization detector (FID), and electron capture detector (ECD).
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)01:20

¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)

When proton-coupled carbon-13 spectra are simplified by a broadband proton decoupling technique, structural information about the coupled protons is lost. Distortionless enhancement by polarization transfer (DEPT) is a technique that provides information on the number of hydrogens attached to each carbon in a molecule. While the DEPT experiment utilizes complex pulse sequences, the pulse delay and flip angle are specifically manipulated. The resulting signals have different phases depending on...
Gas Chromatography: Types of Columns and Stationary Phases01:17

Gas Chromatography: Types of Columns and Stationary Phases

Gas chromatography (GC) relies on stationary phases to separate and analyze components in a sample. There are two main types of stationary phases: liquid and solid. Liquid stationary phases are non-volatile, thermally stable, and chemically inert liquids coated onto the column. Solid stationary phases are particles of adsorbent material, such as silica gel or molecular sieves.
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Multilayer coatings for a multichannel Wolter-like x-ray imager with large field of view and high resolution: Experimental development and study of the instrument performance.

The Review of scientific instruments·2026
Same author

Association between neoadjuvant paclitaxel dose intensity and outcomes in early triple-negative and HER2-positive breast cancer: a real-world data analysis.

ESMO real world data and digital oncology·2026
Same author

Revealing Dentin Multiscale Structures Using High-Resolution Transmission Electron Microscopy.

Journal of dental research·2025
Same author

Hypocretinergic dysfunction in neuromyelitis optica with coma-like episodes.

BMJ case reports·2025
Same author

Enhancing clinical decision support with genomic tools in breast cancer: A Scottish perspective.

Breast (Edinburgh, Scotland)·2024
Same author

Evaluating newly approved drugs in combination regimens for multidrug-resistant tuberculosis with fluoroquinolone resistance (endTB-Q): study protocol for a multi-country randomized controlled trial.

Trials·2023

Related Experiment Video

Updated: Jun 1, 2026

High Resolution Phonon-assisted Quasi-resonance Fluorescence Spectroscopy
10:40

High Resolution Phonon-assisted Quasi-resonance Fluorescence Spectroscopy

Published on: June 28, 2016

Resonant cavity gas-phase polarimeter.

J Poirson1, M Vallet, F Bretenaker

  • 1Laboratoire d'Electronique Quantique-Physique des Lasers, Unité Mixte de Recherche du Centre National de la Recherche Scientifique 6627, Université de Rennes I, Campus de Beaulieu, F-35042 Rennes Cedex, France.

Analytical Chemistry
|June 8, 2011
PubMed
Summary

A novel high-sensitivity polarimeter utilizing Fabry-Perot cavity physics enables precise gas-phase chirality measurements. This advancement achieves a noise level of 10⁻⁶°, significantly improving optical activity detection for chiral molecules.

More Related Videos

Hyperpolarized Xenon for NMR and MRI Applications
16:20

Hyperpolarized Xenon for NMR and MRI Applications

Published on: September 6, 2012

Related Experiment Videos

Last Updated: Jun 1, 2026

High Resolution Phonon-assisted Quasi-resonance Fluorescence Spectroscopy
10:40

High Resolution Phonon-assisted Quasi-resonance Fluorescence Spectroscopy

Published on: June 28, 2016

Hyperpolarized Xenon for NMR and MRI Applications
16:20

Hyperpolarized Xenon for NMR and MRI Applications

Published on: September 6, 2012

Area of Science:

  • Physical Chemistry
  • Analytical Chemistry
  • Spectroscopy

Background:

  • Chirality is a fundamental property in chemistry and biology, crucial for molecular interactions.
  • Gas-phase chirality measurements are challenging due to low concentrations and sensitivity requirements.
  • Traditional polarimeters often lack the sensitivity needed for subtle optical activity detection in the gas phase.

Purpose of the Study:

  • To demonstrate a high-sensitivity polarimeter for gas-phase chirality analysis.
  • To leverage Fabry-Perot cavity physics for enhanced measurement sensitivity.
  • To apply the device for monitoring enantiomeric racemization in the gas phase.

Main Methods:

  • Development of a polarimeter based on the eigenstates of a Fabry-Perot cavity.
  • Measurement of optical rotation induced by vapor-phase (R)-(+)-limonene and (S)-(-)-limonene.
  • Experimental determination of the instrument's noise level.

Main Results:

  • Demonstrated measurement of optical rotations as low as 50 (±1) × 10⁻⁵° for limonene enantiomers.
  • Achieved and experimentally verified a noise level corresponding to a rotation of 10⁻⁶°.
  • Successfully applied the polarimeter to monitor enantiomer mixing and racemization of limonene.

Conclusions:

  • The demonstrated Fabry-Perot cavity-based polarimeter offers significantly improved sensitivity for gas-phase chirality measurements.
  • The device enables precise detection of optical activity and monitoring of enantiomeric processes.
  • This technology has potential applications in chiral analysis and reaction monitoring.