Related Experiment Video
Updated: Jun 1, 2026

Fluorescent Lateral Flow Immunoassay Based on Quantum Dots Nanobeads
Published on: June 28, 2024
Cross-talk-free simultaneous fluoroimmunoassay of two biomarkers based on dual-color quantum dots
Zhijuan Cao1, Huan Li, Choiwan Lau
1School of Pharmacy, Fudan University, Shanghai, PR China.
Abstract:
In this article, we demonstrate the fabrication and simultaneous fluorescent detection of two biomarkers related to lung cancer. Polystyrene microspheres (PSM) were introduced as biomolecular immobilizing carriers and a 96-well filter plate was used as the separation platform. The whole experiment could be effectively carried out in a homogeneous system, as exemplified by the detection of carcinoembryonic antigen (CEA) and neuron specific enolase (NSE). First, two capture antibodies for CEA and NSE were immobilized on the PSM surface. Next, they reacted successively with two antigens and two modified detection antibodies. Finally, these two biomarkers could be recognized by streptavidin-conjugated quantum dots (QD) and goat-anti-FITC conjugated QD with a detection limit of 0.625 ng mL(-1), which was lower than the clinical cut-off level. The protocol showed good precision within 6.36% and good recovery in the range of 90.86-105.02%. Compared with several other assay formats reported previously, our new technique is competitive or even better. Furthermore, the immunosensor was successfully illustrated in 20 serum samples. Overall, this new immunoassay offers a promising alternative for the detection of biomarkers related to cancer diseases, taking advantage of simplicity, specificity, sensitivity and cost-efficiency.