Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Reproducibility and inter-vendor variability of left ventricular deformation measurements by three-dimensional
Etienne Gayat1, Homaa Ahmad, Lynn Weinert
1University of Chicago Medical Center, Chicago, Illinois 60637, USA.
Background:
Myocardial deformation measurements using two-dimensional speckle-tracking echocardiography (STE) are known to vary among vendors. The intervendor agreement of three-dimensional (3D) deformation indices has not been studied. The goals of this study were to determine the intervendor agreement of 3D STE-based measurements of left ventricular (LV) deformation parameters to investigate the intrinsic variability of these measurements and identify the sources of intervendor differences.
Methods:
Real-time full-volume images obtained in 30 subjects with normal LV systolic function using two vendors' equipment (V1 and V2) on the same day were analyzed by two independent observers using two software packages (S1 and S2). Agreement between three technique combinations (V1/S1, V2/S2, and V1/S2) and their intrinsic reproducibility (interobserver and intraobserver agreement) were assessed using intraclass correlation coefficients. Parameters of LV deformation included global longitudinal strain, twist, 3D displacement, and 3D strain and its radial, longitudinal, and circumferential components.
Results:
For all three combinations, intertechnique agreement was poor (intraclass correlation coefficient < 0.4), always beyond the intrinsic variability. For all measured parameters, the intertechnique agreement was better when the same software package was used with images from different vendors (V2/S2 vs V1/S2) than when images from same vendor were analyzed using different software (V1/S2 vs V1/S1).
Conclusions:
Three-dimensional STE-derived LV deformation parameters are highly vendor dependent, and the discordance levels are beyond intrinsic measurement variability of any of the tested combinations of imaging equipment and analysis software. This intervendor discordance must be taken into account when interpreting 3D deformation data.
More Related Videos
06:34Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...