Cochlear reimplantation with same device: surgical and audiologic results
Kim S van der Marel1, Jeroen J Briaire, Berit M Verbist
1Department of Otorhinolaryngology, Leiden University Medical Center, Leiden, the Netherlands.
The Laryngoscope
|June 8, 2011
Summary
Cochlear reimplantation with the same device allows for accurate electrode array positioning and rapid recovery of speech perception. Performance can be regained or even improved within weeks after cochlear implant reimplantation.
Area of Science:
- Otolaryngology
- Neurosurgery
- Biomedical Engineering
Background:
- Cochlear implantation is a common treatment for severe to profound hearing loss.
- Device failure necessitates cochlear reimplantation, raising concerns about electrode array placement and hearing outcomes.
Purpose of the Study:
- To evaluate the feasibility of achieving identical electrode array insertion depths during cochlear reimplantation.
- To assess the maintenance and recovery of speech perception performance after cochlear reimplantation using the same device.
Main Methods:
- Retrospective case series of 15 patients (12 adults, 3 children) undergoing cochlear reimplantation with a HiRes90K device and HiFocus 1J electrode.
- Multislice computed tomography (CT) scans were used to compare the positions of original and reimplanted electrode arrays.
- Speech perception scores (consonant-vowel-consonant test) were compared before and after reimplantation.
Main Results:
- All reimplantations were successful with no complications reported.
- Postoperative CT scans showed an average electrode array displacement of only 0.59 mm.
- Immediate open-set speech understanding was achieved, with speech perception returning to pre-failure levels within weeks and significantly improving at 3 months.
Conclusions:
- Cochlear reimplantation with the same device enables accurate replication of electrode array position.
- Speech perception can be successfully regained and even enhanced shortly after cochlear reimplantation.