Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Image quality in reduced-dose coronary CT angiography
Nina V Gagarina1, Roy Irwan, Galina Gordina
1First Moscow State Medical University, Department of Radiology, Moscow, Russia. ninagagarina@rambler.ru
Reduced-dose coronary computed tomography angiography (CCTA) significantly lowers radiation exposure by 80% without compromising image quality. This low-dose CCTA protocol maintains diagnostic accuracy for coronary artery disease assessment.
Area of Science:
- Cardiovascular Imaging
- Radiology
- Medical Physics
Background:
- Growing concerns over radiation-induced cancer risk necessitate reduced-dose protocols for coronary computed tomography angiography (CCTA).
- Previous comparisons of full-dose and reduced-dose CCTA were inconclusive due to differing patient cohorts.
Purpose of the Study:
- To compare the diagnostic performance and image quality of a new low-dose CCTA protocol against a standard full-dose protocol.
- To evaluate the impact of an 80% radiation dose reduction on CCTA image quality and diagnostic accuracy.
Main Methods:
- A prospective electrocardiogram-gated acquisition was performed on 70 patients with intermediate coronary artery disease (CAD) probability using both standard and low-dose CCTA protocols.
- The low-dose protocol utilized an increased phase window (75% of R-R interval) compared to the standard protocol (10% of R-R interval) on a 320-detector row CT scanner.
- Tube voltages of 120/100 kV were used for both protocols.
Main Results:
- Despite an 80% mean dose reduction (4.9 to 0.98 mSv), visual image quality was not significantly affected.
- Contrast level, image noise, and contrast-to-noise ratio (CNR) were similar between the standard and low-dose protocols.
- A strong correlation between image quality and heart rate variability was observed with the low-dose protocol (r = -0.58, P = .01), unlike the full-dose protocol (r = -0.07, P = .77).
Conclusions:
- An 80% dose reduction in CCTA using a prospective electrocardiogram-gated acquisition generally does not compromise image quality of the coronary vasculature.
- The enhanced performance of the low-dose protocol is attributed to an increased phase window, which improves image quality at the expense of increased sensitivity to heart rate variability.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...