Multiple immunoassay systems are negatively interfered by circulating cardiac troponin I autoantibodies
Gusheng Tang1, Yu Wu, Weiguo Zhao
1Department of Laboratory Medicine, Changhai Hospital, Second Military Medical University, Shanghai, China.
Circulating cardiac troponin I (cTnI) autoantibodies have recently been detected in more and more patients with myocardial injury. In the present study, a total of 121 patients with acute myocardial infarction (AMI) were screened for cTnI autoantibodies using an indirect ELISA. Positive results were further confirmed by Western blot analysis. As a result, 13 autoantibody-positive sera were identified, in which cTnI values detected by different immunoassay systems are very different. Further evaluation revealed low recovery in one of the 13 samples with the Access 2 system (Beckman Coulter, 2^ generation), one low and one moderate recovery sample with Architect i2000 (Abbott), one low and two moderate with AxSYM (Abbott), two low and three moderate with Dimension Xpand (Dade Behring, 2^ generation), and four low and one moderate with Vidas (bioMérieux). Our work demonstrates that circulating cTnI autoantibodies occur in part of patients with AMI and, for the first time to our knowledge, shows that these autoantibodies can result in considerable negative interference in all the five commonly used cTnI immunoassay systems, which may lead to incorrect diagnoses and following treatments. The indirect ELISA established in our laboratory is suitable for a rapid preliminary screening for cTnI autoantibody in clinical work.
Circulating cardiac troponin I (cTnI) autoantibodies have recently been detected in more and more patients with myocardial injury. In the present study, a total of 121 patients with acute myocardial infarction (AMI) were screened for cTnI autoantibodies using an indirect ELISA. Positive results were further confirmed by Western blot analysis. As a result, 13 autoantibody-positive sera were identified, in which cTnI values detected by different immunoassay systems are very different. Further evaluation revealed low recovery in one of the 13 samples with the Access 2 system (Beckman Coulter, 2^ generation), one low and one moderate recovery sample with Architect i2000 (Abbott), one low and two moderate with AxSYM (Abbott), two low and three moderate with Dimension Xpand (Dade Behring, 2^ generation), and four low and one moderate with Vidas (bioMérieux). Our work demonstrates that circulating cTnI autoantibodies occur in part of patients with AMI and, for the first time to our knowledge, shows that these autoantibodies can result in considerable negative interference in all the five commonly used cTnI immunoassay systems, which may lead to incorrect diagnoses and following treatments. The indirect ELISA established in our laboratory is suitable for a rapid preliminary screening for cTnI autoantibody in clinical work.
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Myocarditis II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Acute Coronary Syndrome I: Introduction
Myocarditis I: Introduction