Related Experiment Video
Updated: Jun 1, 2026

Fast Enzymatic Processing of Proteins for MS Detection with a Flow-through Microreactor
Published on: April 6, 2016
Membrane-based emitter for coupling microfluidics with ultrasensitive nanoelectrospray ionization-mass spectrometry
Xuefei Sun1, Ryan T Kelly, Keqi Tang
1Biological Sciences Division, Pacific Northwest National Laboratory, Richland, Washington 99352, United States.
None:
An integrated poly(dimethylsiloxane) (PDMS) membrane-based microfluidic emitter for high-performance nanoelectrospray ionization mass spectrometry has been fabricated and evaluated. The ∼100-microm-thick emitter was created by cutting a PDMS membrane that protrudes beyond the bulk substrate. The reduced surface area at the emitter enhances the electric field and reduces wetting of the surface by the electrospray solvent. As such, the emitter enables highly stable electrosprays at flow rates as low as 10 nL/min and is compatible with electrospray solvents containing a large organic component (e.g., 90% methanol). This approach enables facile emitter construction and provides excellent stability, reproducibility, and sensitivity as well as compatibility with multilayer soft lithography.
More Related Videos
07:55Integrated Cell Manipulation Platform Coupled with the Single-probe for Mass Spectrometry Analysis of Drugs and Metabolites in Single Suspension Cells
Published on: June 21, 2019
08:22Electrowetting-based Digital Microfluidics Platform for Automated Enzyme-linked Immunosorbent Assay
Published on: February 23, 2020