Related Experiment Video
Updated: Jun 1, 2026

Contrast Enhanced Vessel Imaging using MicroCT
Published on: January 27, 2011
Geometric calibration of a micro-CT system and performance for insect imaging
Zhanli Hu1, Jianbao Gui, Jing Zou
1Paul Lauterbur Center for Biomedical Imaging Research, Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China. zl.hu@siat.ac.cn
Abstract:
Micro-CT with a high spatial resolution in combination with computer-based-reconstruction techniques is considered a powerful tool for morphological study of insects. The quality of CT images crucially depends on the precise knowledge of the scan geometry of the micro-CT system. In this paper, we have proposed a method to calculate the deviation of rotating axis for compensating deficiency of existing methods. A practical application of this geometric calibration method of the micro-CT system for insect imaging is presented. We have performed the computer-simulation study and experimental study with our prototype micro-CT system. The results demonstrate that the proposed technique is accurate and robust. In addition, we have evaluated the imaging characteristics of the detector in terms of modulation-transfer function (MTF). Finally, insect imaging performance and image reconstruction from data acquired with different energies are presented.

