N Bârsan1, J R Stetter, M Findlay
1Institute of Physical and Theoretical Chemistry, University of Tübingen, Auf der Morgenstelle 8, D-72076 Tübingen, Germany, BCPS Department, Illinois Institute of Technology, Chicago, Illinois 60616, and TSI Incorporated, P.O. Box 64394, St. Paul, Minnesota 64394.
Abstract:
A comparison of the stability and sensitivity for two different sensor types (semiconductor SnO(2) devices, amperometric electrochemical sensors) has been performed. Sensitivities and drifts in the signal and in the background for various concentrations of CO have been studied for thick-film SnO(2) sensors (Pt and Pd doped) for a period in excess of 8 months. Similar performance data have been recorded for commercial amperometric sensors for a period in excess of 4 years. The two different sensor types investigated here were also compared to the well-known commercial Figaro TGS 822 sensor at similar concentrations. An objective approach for comparing different types of sensors has been developed using the "analytical sensitivity".
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Amperometry: Overview
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...