Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
A chronoamperometric method to estimate ionophore loss from ion-selective electrode membranes
1Department of Chemistry, Rhodes College, 2000 North Parkway, Memphis, Tennessee 38112.
Abstract:
A novel chronoamperometric method was developed to estimate the concentration of a neutral ionophore in fixed-site, dioctyl sebacate plasticized, poly(vinyl chloride)-based, ion-selective electrode membranes. The membranes contained between 0.5 and 16 mmol/kg valinomycin. The chronoamperometric technique was used to estimate the valinomycin concentration in freshly prepared membranes and after extraction of some of the ionophore from the membranes using dioctyl sebacate. Replicate measurements indicated a relative standard deviation in the calculated valinomycin concentration of less than 10%, and these values accurately represented the true concentration of valinomycin within 10%. The method permitted an estimate of the valinomycin concentration after valinomycin was leached from a membrane. The results of preliminary experiments using heparinized dog blood suggest that blood protein adsorption does not interfere qualitatively or quantitatively with the analysis.
More Related Videos
08:06Merging Ion Concentration Polarization between Juxtaposed Ion Exchange Membranes to Block the Propagation of the Polarization Zone
Published on: February 23, 2017
06:59A Fluorescence-Based Assay of Membrane Potential for High-Throughput Functional Study of Two Endogenous Ion Channels in Two Epithelial Cell Lines
Published on: June 22, 2022
Related Concept Videos
Potentiometry: Membrane Electrodes
Amperometry: Overview
Interfacial Electrochemical Methods: Overview
Ion-Exchange Chromatography
Dialysis
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...