Atmospheric-pressure Molecular Imaging of Biological Tissues and Biofilms by LAESI Mass Spectrometry
Published on: September 3, 2010
Improvements in LC/Electrospray Ion Trap Mass Spectrometry Performance Using an Off-Axis Nebulizer
1Analytical and Chemical Sciences, Research Triangle Institute, P.O. Box 12194, Research Triangle Park, North Carolina 27709.
Abstract:
Charged residues from the electrospray process have been hypothesized to limit the sensitivity and dynamic range of an ion trap mass spectrometry operation. Incorporation of an off-axis nebulizer (positioned 90-95° from the sampling orifice) was found to drastically reduce the detrimental effects caused by the charged particles or droplets compared to typical on-axis nebulization configurations (spraying 10-20° from sampling orifices). The off-axis nebulizer reduced total ion currents that enter the ion trap (through the reduction of charged residues) by a factor of 5-7 while resulting in an increase of analyte [M + H](+) signal by a factor of 6 compared to an on-axis sprayer at flow rates of 20 μL/min. At higher flow rates (e.g., 800 μL/min) these enhancements are more evident. At flows greater than 200 μL/min, off-axis nebulization reduced total ion current that enters the ion trap by a factor of 30 and resulted in a factor of more than 20 increase in [M + H](+) signal relative to on-axis nebulization. Incorporation of the off-axis nebulizer improved the detection limit and precision for determination of dihydroxyvitamin D(3) in plasma compared to on-axis nebulization. The LC/MS/MS detection limits obtained for the off-axis nebulizer on the ion trap was within a factor of 2 from the detection limit determined by the triple quadrupole. The relative standard deviation of the dihydroxyvitamin D(3) determination was less than 8% for both off-axis ion trap and triple-quadrupole determinations.
More Related Videos
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
08:23Real-time Breath Analysis by Using Secondary Nanoelectrospray Ionization Coupled to High Resolution Mass Spectrometry
Published on: March 9, 2018
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Analyzers: Common Types
Mass Analyzers: Overview
Tandem Mass Spectrometry
Mass Spectrometers
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview