Multiplex Cytokine Profiling of Stimulated Mouse Splenocytes Using a Cytometric Bead-based Immunoassay Platform
Published on: November 9, 2017
Analytical performance of a multiplexed, bead-based cytokine detection system in small volume samples
Martin Bomert1, Gabi Köllisch, Marjut Roponen
1Institute of Laboratory Medicine and Pathobiochemistry, Molecular Diagnostics, Philipps University, Marburg, Germany.
Clinical Chemistry and Laboratory Medicine
|June 14, 2011
Summary
The cytometric bead array (CBA) offers a reliable and consistent method for multiplexed cytokine measurement in large studies. This cytokine assay requires less sample volume than traditional ELISA, making it suitable for pediatric asthma and allergy research.
Area of Science:
- Immunology
- Biotechnology
- Clinical Research
Background:
- Multiplexed cytokine measurement offers advantages over traditional ELISA for large-scale studies.
- The cytometric bead array (CBA) is a suspension multiplexed protein array.
- Reliability and consistency of CBA in longitudinal studies needed evaluation.
Purpose of the Study:
- To define the reliability and consistency of the cytometric bead array (CBA) in large-scale, longitudinal studies.
- To evaluate CBA performance for measuring specific cytokines in pediatric samples.
- To compare CBA analytical performance against conventional ELISA.
Main Methods:
- Measured interleukin (IL)-5, IL-10, tumor necrosis factor-α (TNF-α), and interferon-γ (IFN-γ) using CBA in pediatric samples.
- Determined analytical performance of CBA, including imprecision, linearity, and recovery.
- Compared CBA results to conventional ELISA.
Main Results:
- Within-run and total imprecision ranged from 5.2%-13.2% for all cytokines.
- Dilution linearity slopes were between 1.01 and 1.31; recovery rates were 97%-113%.
- CBA showed good correlation (r=0.82-0.93) with ELISA and required 25% of the sample volume.
Conclusions:
- CBA demonstrates high reproducibility and sensitivity for cytokine measurement.
- The CBA format is well-suited for small sample volumes.
- CBA is a reliable and consistent method for large-scale, longitudinal studies.