A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
[About an implantable cardioverter-fibrillator]
M-P Douchet1, B Bakouboula, F Bronner
1Service de cardiologie, hôpitaux universitaires de Strasbourg, 1, place de l'Hôpital, 67091 Strasbourg, France. Marie-Pierre.Douchet@chru-strasbourg.fr
Abstract:
A 61-year-old man has been implanted with a Ventritex Profile MD V-186 HV3 ICD for ischemic cardiomyopathy with sustained inducible VT. Three years later, this patient received several inappropriate shocks during the device's interrogation. These shocks provoked ventricular fibrillation. They were caused by a failing soldering between the system random accessory memory (SRAM) module and the hybrid circuit of the device. The device was explanted in emergency.
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiac Catheterization IV: Nursing Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...