Abstract:
Cloth-covered Starr-Edwards caged ball valves implanted in the aortic and mitral valve positions for 39 years were extracted. Both showed valve dysfunction resulting from pannus overgrowth. The metal cages of the Starr-Edwards valves were covered with worn cloth. This case indicates the extended durability of Starr-Edwards valves and the importance of the design and materials of prosthetic heart valves to avoid pannus overgrowth and prosthetic valve abrasion.
Related Concept Videos
Mitral Stenosis III: Medical Management
Mitral Stenosis I: Introduction
Mitral Regurgitation III: Medical Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...