High flow through a nasal cannula and CPAP effect in a simulated infant model
Teresa A Volsko1, Kathy Fedor, Jason Amadei
1Children's Hospital Medical Center of Akron, Akron, OH, USA. tvolsko@chmca.org
Background:
Limited data are available to describe the CPAP effects that can be expected when using high flow with a traditional nasal cannula.
Objective:
To describe the relationship between the pressure generated at the airway opening and flow through a nasal cannula using a simulated infant model. We hypothesized that positive pressure generated by a standard cannula at flows > 2 L/min would be minimal and clinically unimportant.
Methods:
Nares were simulated with holes drilled in a plastic fixture. A nares template for CPAP prongs served as a sizing template for the holes. Small, medium, and large nares fixtures were constructed and connected to a lung simulator that simulated spontaneous breathing. Respiratory muscle pressure was simulated by setting a waveform and adjusting the amplitude to deliver a range of tidal volumes (V(T)) from 3 mL to 12 mL. Lung compliance and resistance were set at 0.5 mL/cm H(2)O and 125 cm H(2)O/L/s, respectively. Nasal cannulas were inserted in the model nares. We assured that the prong occlusion of the nares did not exceed 50%. Cannula flow was adjusted from 2-6 L/min in 1-L/min increments. Data were averaged over 20 breaths. Mean airway pressure and percent change in V(T) were recorded.
Results:
The greatest effect on V(T) (mean ± SD 0.16 ± 0.10 mL) and pressure change (mean ± SD 0.7 ± 0.5 cm H(2)O) occurred with the premature cannula. The least effect on pressure (mean ± SD 0.3 ± 0.22 cm H(2)O) and V(T) change (mean ± SD 0.01 ± 0.02 mL) occurred with the infant cannula.
Conclusions:
Clinically important pressures were not generated by high flows with a standard nasal cannula. The differences in spontaneous V(T) across all flows were negligible.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management