Polyimide polymer glass-free capillary columns for gas chromatography
Jackie G Webster1, Susan S Marine, Neil D Danielson
1Levey Hall, 4200 North University Blvd., Miami University Middletown, Middletown, OH 45042, USA.
Flexible polyimide capillary columns offer a glass-free alternative for gas chromatography (GC). Coated polyimide columns successfully separate volatile organics, showing potential for portable GC instruments.
Area of Science:
- Analytical Chemistry
- Chromatography
- Materials Science
Background:
- Traditional fused silica capillary columns are widely used in gas chromatography (GC).
- Development of flexible, glass-free alternatives is desirable for specialized applications like portable GC.
- Polymeric polyimide tubing presents a potential candidate for GC capillary columns.
Purpose of the Study:
- To evaluate polyimide capillary tubing as a substrate for gas chromatography (GC) columns.
- To investigate the performance of polyimide columns coated with phenylmethylsilicone (PMS) stationary phase.
- To assess the suitability of these columns for analyzing volatile organic compounds.
Main Methods:
- Polyimide capillary tubing (0.32 mm i.d. × 3 m) was cleaned and annealed.
- Columns were coated with phenylmethylsilicone (PMS) using the static method.
- Gas chromatography with a flame ionization detector and split injection was employed for analysis.
- Separations were performed isothermally using volatile organic compounds and a Grob test mix.
Main Results:
- Coated polyimide columns successfully separated alkanes, amines, and fatty acid methyl esters.
- Relative standard deviations for retention time and peak area were 0.5% and 2.5%, respectively.
- A 3 m PMS-coated column with a retention gap achieved a plate count of 3200 (plate height of 1 mm).
- While exhibiting some peak tailing compared to fused silica, headspace analyses were viable.
Conclusions:
- Polyimide capillary columns, particularly when coated with PMS, are suitable for separating volatile organics in GC.
- Their flexibility and small winding diameter make them promising for portable GC applications.
- Despite minor limitations, PMS-coated polyimide columns demonstrate viability for specific analyses like headspace applications.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...