Stent selection and tips on placement technique of EUS-guided biliary drainage: transduodenal and transgastric
Takao Itoi1, Hiroyuki Isayama, Atsushi Sofuni
1Department of Gastroenterology and Hepatology, Tokyo Medical University, 6-7-1 Nishishinjuku, Shinjukuku, Tokyo, Japan. itoi@tokyo-med.ac.jp
Abstract:
Endoscopic ultrasonography-guided biliary drainage (EUS-BD) has been carried out as an alternative to the percutaneous or surgical approach when endoscopic retrograde cholangiopancreatography fails. However, there is no standard technique or device for EUS-BD. In this review, we focus on how we choose the stents and described our tips on this EUS-BD technique. The plastic stent (PS) and the self-expandable metallic stent (SEMS) are used for EUS-BD. The latter is further divided into the fully covered SEMS (FCSEMS), partially covered SEMS (PCSEMS), and uncovered SEMS (UCSEMS) types. Although PS is not expensive, the duration of stent patency is short. SEMS is expensive but the duration of stent patency is long. With UCSEMS, basically there is no stent malpositioning; however, if the gap between the bile duct and the GI tract becomes displaced, bile leakage from the mesh of the stent is likely to occur. Though there is no bile leakage with FCSEMS, the side branch of the bile duct may become occluded, and migration and dislocation sometimes occur. PCSEMS is basically similar to FCSEMS. When EUS-BD was first developed, drainage by PS was common, although reports on drainage by SEMS have increased recently.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Enteral Nutrition II: Nasointestinal and Gastrostomy Feeding
Nasointestinal Feeding
Nasointestinal feeding involves placing a tube through...