Effect of scanner in longitudinal studies of brain volume changes
Hidemasa Takao1, Naoto Hayashi, Kuni Ohtomo
1Department of Radiology, Graduate School of Medicine, University of Tokyo, Tokyo, Japan. takaoh-tky@umin.ac.jp
Purpose:
To evaluate the effects of drift in scanner hardware and inter-scanner variability on global and regional brain volume changes using longitudinal data obtained on two scanners of the exact same model.
Materials And Methods:
A total of 208 subjects (70 females and 138 males, mean age = 56.5 ± 10.1 years) were scanned twice, at an interval of approximately 2 years (mean interval = 1.9 ± 0.15 years), using two 3.0 Tesla (T) scanners of the exact same model. The subjects were divided into four groups according to the combination of scanners used. The effects of scanner drift and inter-scanner variability on global and regional brain volume changes were evaluated using voxel-based morphometry (VBM) and structural image evaluation using normalization of atrophy (SIENA).
Results:
The mean whole brain volume calculated using VBM increased by 17.2 ± 29.3 mL, and the changes were significantly different between groups. The mean percentage brain volume change (PBVC) calculated using SIENA was -0.46 ± 0.71 %, and did not differ between groups. VBM showed several regions with a significant increase or decrease in volume; however, SIENA showed that edge displacement values were negative in most regions.
Conclusion:
Even with scanners of the exact same model, scanner drift, and inter-scanner variability can cancel out actual longitudinal brain volume change. SIENA, which corrects for differences in imaging geometry using the outer skull surface for both time points, can reduce the effects of scanner drift and inter-scanner variability on longitudinal morphometric results.
More Related Videos
Related Concept Videos
Longitudinal Studies
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies IV: Magnetic Resonance Imaging
Longitudinal Research
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography