Comparison of radiation dose and image quality: 320-MDCT versus 64-MDCT coronary angiography
Atif Khan1, Faisal Khosa, Khurram Nasir
1Department of Radiology, Beth Israel Deaconess Medical Center, Boston, MA 02215, USA.
Objective:
The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating.
Materials And Methods:
One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus.
Results:
The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001).
Conclusion:
Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...