Laboratory evaluation of 10 heat and moisture exchangers using simulated aeromedical evacuation conditions
Huda S Suliman1, Stephen E Fecura, Jonathan Baskin
1Hyperion Biotechnology, Inc., 13302 Langtry, San Antonio, TX 78248, USA.
Abstract:
Heat and moisture exchangers (HMEs) are used for airway humidification in mechanically ventilated patients and have been evaluated only under hospital conditions. U.S. Air Force aeromedical evacuation transports are performed under rugged conditions further complicated by the cold and dry environment in military aircrafts, and HMEs are used to provide airway humidification for patients. This study evaluated 10 commercial HMEs using a test system that simulated aeromedical evacuation conditions. Although the American National Standards Institute recommends inspired air to be at an absolute humidity value of > or = 30 mg/L for mechanically ventilated patients, the highest absolute humidity by any HME was approximately 20 mg/L. Although none of the HMEs were able to maintain a temperature high enough to achieve the humidity standard of the American National Standards Institute, the clinical significance of this standard may be less important than the relative humidity maintained in the respired air, especially on evacuation flights of short duration.
More Related Videos
09:29An Air-liquid Interface Bronchial Epithelial Model for Realistic, Repeated Inhalation Exposure to Airborne Particles for Toxicity Testing
Published on: May 13, 2020
09:34Thermal Preconditioning During Ex-vivo Lung Perfusion for the Rehabilitation of Damaged Lung Grafts before Transplantation
Published on: October 31, 2025