A direction-selective flattening filter for clinical photon beams. Monte Carlo evaluation of a new concept
Ndimofor Chofor1, Dietrich Harder, Kay Willborn
1Working Group Medical Radiation Physics, Carl von Ossietzky University Oldenburg, Germany. cmark23@hotmail.com
Abstract:
A new concept for the design of flattening filters applied in the generation of 6 and 15 MV photon beams by clinical linear accelerators is evaluated by Monte Carlo simulation. The beam head of the Siemens Primus accelerator has been taken as the starting point for the study of the conceived beam head modifications. The direction-selective filter (DSF) system developed in this work is midway between the classical flattening filter (FF) by which homogeneous transversal dose profiles have been established, and the flattening filter-free (FFF) design, by which advantages such as increased dose rate and reduced production of leakage photons and photoneutrons per Gy in the irradiated region have been achieved, whereas dose profile flatness was abandoned. The DSF concept is based on the selective attenuation of bremsstrahlung photons depending on their direction of emission from the bremsstrahlung target, accomplished by means of newly designed small conical filters arranged close to the target. This results in the capture of large-angle scattered Compton photons from the filter in the primary collimator. Beam flatness has been obtained up to any field cross section which does not exceed a circle of 15 cm diameter at 100 cm focal distance, such as 10 × 10 cm(2), 4 × 14.5 cm(2) or less. This flatness offers simplicity of dosimetric verifications, online controls and plausibility estimates of the dose to the target volume. The concept can be utilized when the application of small- and medium-sized homogeneous fields is sufficient, e.g. in the treatment of prostate, brain, salivary gland, larynx and pharynx as well as pediatric tumors and for cranial or extracranial stereotactic treatments. Significant dose rate enhancement has been achieved compared with the FF system, with enhancement factors 1.67 (DSF) and 2.08 (FFF) for 6 MV, and 2.54 (DSF) and 3.96 (FFF) for 15 MV. Shortening the delivery time per fraction matters with regard to workflow in a radiotherapy department, patient comfort, reduction of errors due to patient movement and a slight, probably just noticable improvement of the treatment outcome due to radiobiological reasons. In comparison with the FF system, the number of head leakage photons per Gy in the irradiated region has been reduced at 15 MV by factors 1/2.54 (DSF) and 1/3.96 (FFF), and the source strength of photoneutrons was reduced by factors 1/2.81 (DSF) and 1/3.49 (FFF).