[Adverse events during transcranial muscle evoked potential monitoring]

Yoshie Kuwahara1, Shima Taguchi, Masahiko Kuroda

  • 1Department of Anesthesiology and Critical Care, Hiroshima University Hospital, Hiroshima 734-8551.

Masui. the Japanese Journal of Anesthesiology
|June 30, 2011
PubMed
Abstract

Insights

Transcranial muscle evoked potential (TC-MsEP) monitoring aids spinal cord function assessment during surgery. However, the Multipulse D-185 device increased adverse events like tracheal tube issues and bite injuries due to masseter muscle contraction.

Area of Science:

  • Neurosurgery
  • Spinal Cord Monitoring
  • Anesthesiology

Background:

  • Transcranial muscle evoked potential (TC-MsEP) monitoring assesses spinal cord motor tract function to prevent paralysis during spine surgery.
  • Recent advancements have improved TC-MsEP monitoring quality.
  • However, adverse events including tracheal tube damage, tongue biting, and mandibular fractures have been reported.

Observation:

  • A retrospective review of adverse events associated with TC-MsEP monitoring using the Multipulse D-185 device during spine surgery was conducted.
  • The study compared the incidence of tracheal tube transformation between the Multipulse D-185 and other stimulation devices.
  • Two specific adverse events, tongue laceration and teeth dislocation, were noted with the Multipulse D-185.

Findings:

  • The Multipulse D-185 device was associated with a significantly higher rate of tracheal tube transformation (11 of 20 cases) compared to other devices (2 of 19 cases; P=0.02).
  • Adverse events, including tongue laceration and teeth dislocation, were attributed to excessive masseter muscle contraction induced by TC-MsEP monitoring.
  • The incidence of tracheal tube issues was significantly higher with the Multipulse D-185.

Implications:

  • The findings suggest a need for preventative strategies against bite injuries during TC-MsEP monitoring.
  • Development of specific ventilation strategies is crucial to manage potential tube occlusion or stenosis when using the Multipulse D-185.
  • Enhanced safety protocols are required for TC-MsEP monitoring, particularly with the Multipulse D-185 device, to mitigate risks of adverse events.