Diaton tonometry: an assessment of validity and preference against Goldmann tonometry
Mark D Doherty1, Zia I Carrim, Damian P O'Neill
1Department of Ophthalmology, St James's University Hospital, Leeds, UK. highburymark@yahoo.co.uk
Background:
To assess agreement between the Diaton, a new transpalpebral tonometer, and Goldmann applanation tonometry, the accepted gold standard.
Design:
Comparative study of two devices in a hospital setting.
Participants:
Two hundred and fifty-one patients attending the eye casualty and general ophthalmology clinics at St James' University Hospital, Leeds between February and December 2009.
Methods:
Intraocular pressure was measured using Goldmann applanation tonometry and Diaton tonometry by one examining ophthalmologist. Patient preference for either technique was also recorded.
Main Outcome Measures:
Intraocular pressure measured by Diaton was compared with intraocular pressure measured by Goldmann applanation tonometry. Limits of agreement were determined using the Bland-Altman method.
Results:
Two hundred and fifty right eyes underwent both Goldmann applanation tonometry and Diaton tonometry. Mean intraocular pressure was 13.8 ± 3.6 mmHg using Goldmann applanation tonometry and 13.2 ± 4.3 mmHg using Diaton tonometry. Upper and lower limits of agreement were +8.4 mmHg and -9.6 mmHg, respectively. Order of intraocular pressure measurement and positioning did not influence limits of agreement in a clinically significant manner. Overall, more patients expressed preference for Diaton tonometry (40.2%) than Goldmann applanation tonometry (30.3%). Those aged 50 or less were more likely to prefer Diaton tonometry.
Conclusions:
The Diaton tonometer is portable, lightweight, user-friendly and well tolerated by patients. However, it shows poor agreement with Goldmann applanation tonometry, thereby precluding it from being regarded as a substitute in routine clinical practice.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...