Related Experiment Video
Updated: May 31, 2026

Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Feasibility study on bonding quality inspection of microfluidic devices by optical coherence tomography
Shiguang Li1, Zhiguang Xu, Soon Fatt Yoon
1Singapore-MIT Alliance, N3.2-01-36, 65 Nanyang Drive, Singapore. lishiguang@tsinghua.org.cn
Abstract:
This paper reports the feasibility of optical coherence tomography (OCT) technology for inspection of bonding quality of microfluidic devices in manufacturing environments. A compact optical-fiber-based OCT is developed and its measurement performance is characterized. A series of microfluidic devices respectively bonded by adhesive tape, thermal method, and oxygen plasma, are inspected. The defects of geometry deformation and sealing completeness are emphasized during measurements. Based on the inspection results, some discoveries related to the production of microfluidic devices are discussed.