Data processing for noninvasive continuous glucose monitoring with a multisensor device
Martin Mueller1, Mark S Talary, Lisa Falco
1Research & Development Department, Solianis Monitoring AG, Zürich, Switzerland. andreas.caduff@solianis.com
Background:
Impedance spectroscopy has been shown to be a candidate for noninvasive continuous glucose monitoring in humans. However, in addition to glucose, other factors also have effects on impedance characteristics of the skin and underlying tissue.
Method:
Impedance spectra were summarized through a principal component analysis and relevant variables were identified with Akaike's information criterion. In order to model blood glucose, a linear least-squares model was used. A Monte Carlo simulation was applied to examine the effects of personalizing models.
Results:
The principal component analysis was able to identify two major effects in the impedance spectra: a blood glucose-related process and an equilibration process related to moisturization of the skin and underlying tissue. With a global linear least-squares model, a coefficient of determination (R²) of 0.60 was achieved, whereas the personalized model reached an R² of 0.71. The Monte Carlo simulation proved a significant advantage of personalized models over global models.
Conclusion:
A principal component analysis is useful for extracting glucose-related effects in the impedance spectra of human skin. A linear global model based on Solianis Multisensor data yields a good predictive power for blood glucose estimation. However, a personalized linear model still has greater predictive power.