The effect of heart rhythm on patient radiation dose with dual-source cardiac computed tomography
Tust Techasith1, Brian B Ghoshhajra, Quynh A Truong
1Cardiac MR PET CT Program, Department of Radiology and Division of Cardiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA.
Background:
To lower the radiation exposure associated with cardiac CT, it is essential to identify all factors that influence radiation dose.
Objectives:
We explored the effect of heart rhythm during scan acquisition on radiation dose with a 64-slice dual-source cardiac CT.
Methods:
Patient and scan data were collected prospectively in 302 consecutive patients referred for a clinical dual-source cardiac CT. Electrocardiograms recorded during acquisition were interpreted by a cardiologist and categorized as (1) normal sinus rhythm (NSR), (2) premature atrial contraction (PAC) or premature ventricular contraction (PVC), or (3) atrial fibrillation or flutter.
Results:
Of the 302 patients, 227 (75.2%) were in NSR and had no ectopy, 55 (18.2%) had PAC/PVC, and 20 (6.6%) had atrial fibrillation or flutter during the scan. Patients with irregular rhythm (PAC/PVC and atrial fibrillation or flutter) were older than patients with regular rhythm (61.0 vs 54.8 years; P = 0.006). Patients with NSR had the lowest estimated radiation dose, followed by PAC/PVC and atrial fibrillation/flutter (9.4, 14.5, 20.9 mSv; P < 0.001). The difference remained significant after adjustments for differences in examination type, tube current and voltage, scan length, pitch, and use of tube current modulation (9.8, 14.1, 17.9 mSv; P < 0.001). No significant association was observed between heart rhythm and subjective image quality although scans with regular rhythm and no ectopy had higher signal-to-noise and contrast-to-noise ratios (P < 0.01).
Conclusion:
Compared to patients with NSR, patients with atrial fibrillation/flutter had the highest radiation exposure, followed by those with PAC/PVC. Even after adjustment for factors associated with radiation exposure, a significant difference in radiation dose persisted. These findings can be used to identify patients who are more likely to receive higher radiation dose when undergoing cardiac CT and to develop future more-efficient scanner algorithms for use in patients with arrhythmias.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...