Safety of nerve conduction studies in patients with implantable cardioverter-defibrillators
Mirosława Derejko1, Paweł Derejko, Andrzej Przybylski
1Institute of Psychiatry and Neurology, Department of Clinical Neurophysiology, Warsaw, Poland. mderejko@yahoo.com
Objective:
A patient with an implantable cardioverter-defibrillator (ICD) may suffer from neuromuscular disorders and may need to undergo a nerve conduction study (NCS). However, a NCS may be a source of electromagnetic interference (EMI). The aim of the present study was to investigate whether the interference from NCS used in a standardised test protocol affects ICD function.
Methods:
Twenty patients (19 males; mean age of 59.8±9.9 years) with implantable ICDs (eight with integrated and 12 with true bipolar leads), treated with amiodarone and with symptoms suggesting neuropathy were included. NCS were conducted using repetitive stimulation with frequency of 2 Hz and single, rectangular pulses of intensity up to 100 mA. Stimulation was performed in standard sites including proximal sites in the arm.
Results:
The impulses generated NCS were not detected by the ICD, irrespective of the site, rate or stimulus intensity.
Conclusions:
Standardised test protocol for an NCS is safe in patients with an ICD regardless of the leads type.
Significance:
Current guidelines which limitate the NCS in patients with ICD may be the subject of revision.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization IV: Nursing Management
Antiarrhythmic Drugs: Class I Agents as Sodium Channel Blockers
Class 1A Antiarrhythmic Drugs: These drugs work by moderately blocking sodium channels,...
Cardiac Catheterization I: Pre-Procedure Overview