Interobserver agreement for sonograms of breast lesions obtained by an automated breast volume scanner
Jing Zhang1, Xing-Jian Lai, Qing-Li Zhu
1Department of Diagnostic Ultrasound, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuaifuyuan, Dongcheng District, Beijing 100730, China. zhang.jing1029@163.com
Objective:
To evaluate the interobserver agreement of radiologists in the description and final assessment of breast sonograms obtained using an automated breast volume scanner (ABVS) using a unique descriptor of three-dimensional ultrasound (3D US) and the Breast Imaging Reporting and Data System (BI-RADS) US lexicon.
Methods:
From October to December 2010, 208 patients were subjected to an ABVS examination in the supine position, and data were automatically sent to the ABVS workstation. Two radiologists independently evaluated 234 breast masses (148 benign and 86 malignant masses) using a unique descriptor from the 3D US and the BI-RADS US lexicon. The reviewers were blinded to the patient's mammographic images, medical history, and pathologic findings. The interobserver agreement was measured using kappa statistics.
Results:
Substantial agreement was obtained for lesion shape, orientation, margin, echo pattern, posterior acoustic features, calcification and final assessment (κ=0.79, 0.74, 0.76, 0.69, 0.68, 0.71 and 0.70, respectively). Fair agreement was obtained for retraction phenomenon and lesion boundary (κ=0.54 and 0.42, respectively).
Conclusions:
The interobserver agreement for breast sonograms obtained by ABVS is good, especially for lesion shape and margin; however, the interobserver agreement for the retraction phenomenon, which is a unique descriptor of coronal-plane 3D US, needs to be improved.
More Related Videos
12:50Lesion Explorer: A Video-guided, Standardized Protocol for Accurate and Reliable MRI-derived Volumetrics in Alzheimer's Disease and Normal Elderly
Published on: April 14, 2014
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called a...