Measurement of the eddy diffusion term in chromatographic columns. I. Application to the first generation of 4.6mm I.D. monolithic columns
Related Experiment Video
Updated: May 31, 2026

Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Fabrice Gritti1, Georges Guiochon
1Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:
The corrected heights equivalent to a theoretical plate (HETP) of three 4.6mm I.D. monolithic Onyx-C(18) columns (Onyx, Phenomenex, Torrance, CA) of different lengths (2.5, 5, and 10 cm) are reported for retained (toluene, naphthalene) and non-retained (uracil, caffeine) small molecules. The moments of the peak profiles were measured according to the accurate numerical integration method. Correction for the extra-column contributions was systematically applied. The peak parking method was used in order to measure the bulk diffusion coefficients of the sample molecules, their longitudinal diffusion terms, and the eddy diffusion term of the three monolithic columns. The experimental results demonstrate that the maximum efficiency was 60,000 plates/m for retained compounds. The column length has a large impact on the plate height of non-retained species. These observations were unambiguously explained by a large trans-column eddy diffusion term in the van Deemter HETP equation. This large trans-rod eddy diffusion term is due to the combination of a large trans-rod velocity bias (≃3%), a small radial dispersion coefficient in silica monolithic columns, and a poorly designed distribution and collection of the sample streamlets at the inlet and outlet of the monolithic rod. Improving the performance of large I.D. monolithic columns will require (1) a detailed knowledge of the actual flow distribution across and along these monolithic rod and (2) the design of appropriate inlet and outlet distributors designed to minimize the nefarious impact of the radial flow heterogeneity on band broadening.
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Principles Of Column Chromatography
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...