Accuracy of pulse oximeter readings from probe placement on newborn wrist and ankle
N Phattraprayoon1, S Sardesai, M Durand
1Department of Pediatrics, USC Division of Neonatal Medicine, LAC+USC Medical Center and Children's Hospital Los Angeles, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA.
Objective:
To compare the accuracy of pulse oximetry oxygen saturation (SpO(2)) measured on the wrist compared with the ipsilateral palm, and SpO(2) measured on the ankle compared with the ipsilateral sole.
Study Design:
In this prospective observational study, neonates admitted to the neonatal intensive care unit were enrolled. We recorded SpO(2) (Masimo Radical-7 pulse oximeter) detected at the palm and ipsilateral wrist initially, then at 30 s, and at 1 min, and we repeated the same procedure over the sole and ipsilateral ankle. We recorded the time to obtain the SpO(2) readings from all these sites. Regression analysis was performed to determine the relationship between paired SpO(2) measurements. The mean difference (bias) and standard deviation of the paired SpO(2) differences (precision) were calculated (Bland-Altman plots).
Result:
A total of 150 patients (birth weight 2381±1020 g, gestational age 34.3±4.3 weeks, median postnatal age 3.5 days (25th-75th percentile 1-16 days)) were enrolled. There was a good correlation between SpO(2) measured at the palm versus the wrist (r=0.95, P<0.001 (right); r=0.97, P< 0.001 (left)) and between SpO(2) measured at the sole versus the ankle (r=0.92, P<0.001 (right); r=0.91, P<0.001 (left)). There was also a good agreement between paired SpO(2) measurements from these sites. The bias and precision for SpO(2) at the right palm and right wrist was 0.08±0.94% and for the left palm and left wrist 0.22±0.87%. Similarly, the bias and precision for SpO(2) at the right sole and right ankle was -0.03±0.93% and for the left sole and left ankle was -0.01±0.93%.
Conclusion:
Our results show that the wrist and ankle can be used as alternative sites to measure SpO(2) in newborn infants in place of the routinely used palm or sole.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement