Background:
The purpose of this study was to present a radiofrequency (RF) endovascular occlusion device (ie, Habib VesCoag Catheter; EMcision Ltd, London, UK) and to evaluate safety and efficacy of the device for complete occlusion of normal porcine vessels.
Methods:
The study included 20 pigs. In each pig, a segmental branch of the right hepatic artery, a branch of the splenic artery, and a branch of one of the renal arteries were catheterized. A single or multiple applications of RF energy were performed until vessel occlusion was achieved. Fifteen days later, angiography was repeated to assess vessel patency. The vessels were then excised for pathological analysis.
Results:
Vessels 2.5 to 6 mm in diameter were treated. Complete occlusion with a single attempt was achieved using a mean amount of energy of 110.67 J in vessels 2.5 to 3 mm, 111.67 J in vessels 3.5 to 4 mm, and 116.63 J in vessels 5 to 6 mm in diameter and was confirmed by angiography at the 15-day follow-up.
Conclusions:
Vascular occlusion can be effectively and safely achieved by endovascular application of RF energy to normal porcine arteries using the Habib VesCoag catheter.