Novel analysis of clinically relevant diagnostic errors in point-of-care devices
K M Shermock1, M B Streiff, B L Pinto
1Department of Pharmacy, Johns Hopkins Medicine, Baltimore, MD 21287, USA. kenneth@jhmi.edu
Background:
To ensure proper clinical decision-making and avoid preventable harm, the quality of point-of-care (POC) device measures is routinely assessed. Traditional analyses may not reveal clinically important diagnostic errors.
Objectives:
To compare results between a novel analytic framework and traditional analyses.
Methods:
Patients in four anticoagulation clinics provided two measures of the International Normalized Ratio (INR) at the same visit as part of routine quality assurance: one via a venous sample and one fingerstick. These were assessed with Hemochron POC devices. Traditional, quarterly, quality assurance assessments emphasized correlation analysis. The novel analysis used enhanced graphics and a validated assessment of clinical decision-making.
Results:
1518 paired INRs were analyzed. The correlation between the POC and laboratory assessments ranged between 0.84 and 0.91. Traditional quality assurance showed that the Hemochron devices were acceptable for continued use in each quarterly analysis. Enhanced graphical analysis demonstrated that the Hemochron devices never reported seven common INR values. The Hemochron devices systematically inflated values < 3 and deflated values > 4, biasing results towards the target INR range. Consequently, the Hemochron devices lead to a different clinical decision than the clinical laboratory measure in 31% of cases (458/1466; 95% confidence interval [CI] 29-34). When the reference INR was low, the Hemochron devices would not result in appropriate dose increases in 52% of cases (95% CI 48-56), placing these patients at risk for a significant adverse drug event.
Conclusions:
Our novel, clinically relevant analysis revealed previously undetected deficiencies in our POC INR devices, and our approach should be adopted by industry, regulators, and institutions.
Related Concept Videos
Documentation of Nursing Diagnosis
In some settings, data-driven computerized decision support systems are in place, allowing for more accurate nursing diagnoses. The database within one of these systems includes diagnostic labels defining characteristics, activities, and indicators for nursing. A nurse enters assessment...
Errors occurring during blood pressure monitoring
Several factors...
Automated Microbial Diagnostics
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...