Effect of tissue mechanical properties on cuff-based blood pressure measurements
H Lan1, A M Al-Jumaily, A Lowe
1Institute of Biomedical Technologies, Auckland University of Technology, Private Bag 92006, Auckland, New Zealand.
Abstract:
This paper presents a 3D finite element upper arm model, validated by experiments as well as clinical data, used to study the error introduced in blood pressure measurements due to variability of arm tissue mechanical properties. The model consists of three separate cylindrical parts: soft tissue, bone and brachial artery. The artery volume changes under the cuff are used to represent the cuff pressure oscillations for analyzing blood pressure measurements. These oscillation trends are identical to observed clinical data. Also an upper arm simulator is designed and built for model validation. The model shows that the variation of soft tissue compressibility introduces an error up to 5% in blood pressure measurements. It is also revealed that the variation of the brachial artery and arm tissue stiffness has an insignificant effect on oscillometric blood pressure measurement method.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...