Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
LV thrombus detection by routine echocardiography: insights into performance characteristics using delayed
Jonathan W Weinsaft1, Han W Kim, Anna Lisa Crowley
1Duke Cardiovascular Magnetic Resonance Center, Duke University Medical Center, Durham, North Carolina 27710, USA.
Objectives:
This study sought to evaluate performance characteristics of routine echo for left ventricular thrombus (LVT).
Background:
Although the utility of dedicated echocardiography (echo) for LVT is established, echo is widely used as a general test for which LVT is rarely the primary indication. We used delayed-enhancement cardiac magnetic resonance (DE-CMR) as a reference to evaluate LVT detection by routine echo.
Methods:
Dedicated LVT assessment using DE-CMR was prospectively performed in patients with left ventricular systolic dysfunction. Echoes were done as part of routine clinical care. Echo and CMR were independently read for LVT and related indexes of LVT size, shape, and image quality/diagnostic confidence. Follow-up was done for embolic events and pathology validation of LVT.
Results:
In this study, 243 patients had routine clinical echo and dedicated CMR within 1 week without intervening events. Follow-up supported DE-CMR as a reference standard, with >5-fold difference in endpoints between patients with versus without LVT by DE-CMR (p = 0.02). LVT prevalence was 10% by DE-CMR. Echo contrast was used in 4% of patients. Echo sensitivity and specificity were 33% and 91%, with positive and negative predictive values of 29% and 93%. Among patients with possible LVT as the clinical indication for echo, sensitivity and positive predictive value were markedly higher (60%, 75%). Regarding sensitivity, echo performance related to LVT morphology and mirrored cine-CMR, with protuberant thrombus typically missed when small (p ≤ 0.02). There was also a strong trend to miss mural thrombus irrespective of size (p = 0.06). Concerning positive predictive value, echo performance related to image quality, with lower diagnostic confidence scores for echoes read positive for LVT in discordance with DE-CMR compared with echoes concordant with DE-CMR (p < 0.02).
Conclusions:
Routine echo with rare contrast use can yield misleading results concerning LVT. Echo performance is improved when large protuberant thrombus is present and when the clinical indication is specifically for LVT assessment.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT