Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
The orientation of orthopaedic metallic devices relative to the frequency-encoding gradient affects susceptibility
Naohide Takeuchi1, Hiromichi Mitsuyasu, Tomonori Nakanishi
1Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kyushu University, Fukuoka.
Purpose:
To evaluate the magnetic susceptibility artifacts associated with different frequency-encoding gradient directions for an angled cephalomedullary device of the proximal femur, and to determine the optimal extremity positioning for reducing artifacts using 0.4 T open MR imaging.
Materials And Methods:
Two different angular devices made of titanium alloy and stainless steel were used. The images were obtained with the frequency-encoding gradient parallel to the rod (Group R) and parallel to the lag screw (Group L). The device positioning was altered in order to obtain images with frequency-encoding gradient parallel to the rod and parallel to the lag screw. The artifact areas associated with the whole device and the lag screw were statistically evaluated.
Results:
For both devices, the mean artifact area in Group L was significantly larger than that in Group R (p<0.05). However, the mean artifact area of the lag screw only in Group L was significantly smaller than that in Group R (p<0.05).
Conclusion:
Susceptibility artifacts for angled cephalomedullary devices can be minimized when the frequency-encoding gradient is parallel to the long axis of the regions of interest. Open MR imaging enables us to obtain the optimal orientation for minimizing susceptibility artifacts.