Body size-adapted dose of contrast material and scanning protocol in 320-detector row CT coronary angiography
Fuminari Tatsugami1, Mitsuru Matsuki, Go Nakai
1Department of Radiology, Osaka Medical College, Takatsuki City, Osaka, Japan. sa104@rg8.so-net.ne.jp
Objective:
Because an increase in patient body size produces lower vessel attenuation and higher image noise in computed tomography coronary angiography (CTCA), a protocol in which the dose of contrast material was adapted to the body weight, and tube voltage and current were adapted to the body mass index (BMI) was evaluated.
Methods:
A total of 136 patients who underwent CTCA were assigned to 1 of 2 protocols; the patients received 40 mL of contrast material (A; n = 52), or 0.7 mL/kg of contrast material (B; n = 84). Tube voltage and current were adapted to the BMI.
Results:
In group A, there was no significant correlation between BMI and image noise (r = -0.21, P = 0.15). However, BMI correlated inversely with contrast-to-noise ratio (CNR; right coronary artery, r = -0.29, P < 0.05; left main coronary artery, r = -0.33, P < 0.05) and image quality score (r = -0.55, P < 0.001). In group B, there was no significant correlation between BMI and image noise (r = -0.14, P = 0.21), CNR (right coronary artery: r = -0.09, P = 0.45; left main coronary artery: r = -0.07, P = 0.55), and image quality score (r = 0.03, P = 0.79).
Conclusions:
Use of a body size-adapted dose of contrast material and scanning protocol results in similar CNR and image quality independent of individual BMI.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...