A dose index as a tool to estimate paediatric patient doses in digital projection radiography
Paula Garcia Castañon1, Maria Luisa España Lopez, Virginia Fernandez Bedoya
1Medical Physics and Radiation Protection Department, Hospital Universitario La Princesa, Diego de Leon 62, 28006 Madrid, Spain. pgarciac.hlpr@salud.madrid.org
Insights
Paediatric patient dose monitoring is crucial due to higher lifetime radiation risks in children. A DICOM header dose index accurately estimates entrance surface dose (ESD) for paediatric chest X-rays, aiding in dose management.
Area of Science:
- Medical Imaging
- Radiology
- Radiation Dosimetry
Background:
- Childhood radiation exposure poses a significantly higher lifetime risk compared to adult exposure.
- Effective paediatric patient dose monitoring is a critical concern in diagnostic radiology.
- Accurate estimation of radiation dose is essential for patient safety and epidemiological studies.
Purpose of the Study:
- To validate the utility of a dose index found in the Digital Imaging and Communication in Medicine (DICOM) header.
- To assess if this dose index can reliably estimate the Entrance Surface Dose (ESD) for paediatric radiological procedures.
- To determine the accuracy of the DICOM dose index for paediatric chest examinations.
Main Methods:
- A study involving 156 paediatric patients undergoing chest examinations.
- Utilized the National Radiation Protection Board Report 318 model for patient age categorization and thickness estimation.
- Compared corrected DICOM dose index values with calculated ESD derived from tube output and radiographic techniques.
Main Results:
- The deviation between the corrected dose index and calculated Entrance Surface Dose (ESD) was within 2% across all age groups studied.
- An exception was noted for patients up to one year of age, where the deviation was slightly higher.
- The DICOM dose index demonstrated high validity as a dosimetric quantity for estimating ESD in paediatric patients.
Conclusions:
- The DICOM header dose index is a valid and reliable tool for estimating Entrance Surface Dose (ESD) in paediatric patients.
- This validated index facilitates the management and expansion of patient dose databases for improved radiation safety.
- The findings support the routine use of DICOM dose index for dose assessment in paediatric radiology.
Abstract:
Radiation exposure during childhood is estimated to have a lifetime risk up to seven times greater than exposures at adult age. Therefore, paediatric patient dose monitoring is a major concern in radiology. The aim of this study is to evaluate the validity of a dose index, displayed at the digital imaging and communication in medicine (DICOM) header of the images, as a dosimetric quantity to estimate the entrance surface dose (ESD) for each procedure. It was carried out over a sample of 156 paediatric patients who underwent a chest examination. National Radiation Protection Board Report 318 model was used to categorise patients by age and to estimate their thickness. Corrected dose index values were compared with calculated ESD, estimated from tube output and radiographic technique. The deviation between both values remained within 2% for every age group, except for patients up to 1 y. Therefore, the index could be used to estimate ESD, allowing to manage greater patient dose databases.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Drug Dosing: Infants and Children
X-ray Imaging