Temperature dependence of the dose response for a solid-state radiochromic dosimeter during irradiation and storage
Peter S Skyt1, Peter Balling, Jørgen B B Petersen
1Department of Medical Physics, Aarhus University Hospital, DK-8000 Aarhus C, Denmark. skyt@phys.au.dk
Purpose:
The dose response of radiochromic dosimeters is based on radiation-induced chemical reactions and is thus likely to be thermally influenced. In this study we have therefore investigated the temperature dependence of the dose response for such dosimeters, regarding both irradiation and storage conditions.
Methods:
Dosimeter samples in cuvettes were irradiated to 5 Gy. The temperature for the different cuvettes during irradiation and post-irradiation storage was varied in the range of 3-30 degrees C in order to quantify the temperature dependence of the dosimeter response. The optical properties of the dosimeter samples were measured using a spectrophotometer before irradiation as well as at several times after irradiation to quantify the temporal variation of dose response (expressed as the optical density change induced by irradiation) as a function of storage temperature.
Results:
The measurements show considerable temperature dependencies of dose response both during irradiation and storage. Fit to an Arrhenius equation revealed an activation energy of 1.4 +/- 0.2 eV for the variation in irradiation temperature, indicating a contribution from a thermally activated process. Variation in dose response at different storage temperatures showed an exponential increase with time followed by a decrease in optical density. Exponential Arrhenius fits to rate constants gave activation energies of 1.7 +/- 0.2 eV for the increase in dose response and 2.3 +/- 0.5 eV for the subsequent decrease, in this case dominated by thermally activated processes.
Conclusions:
Due to the exponential dependencies, stabilization of the dosimeter during irradiation at low temperatures (e.g., 5 degrees C) is preferable in clinical use to optimize the accuracy of the dose response. In addition, a low storage temperature is recommended in order to minimize the post-irradiation temporal change in dose response and thereby increase the post-irradiation stability of the dosimeter. The measurements in this study show that if the observed temperature and temporal dependencies are not considered, this could potentially deteriorate the accuracy of the dosimeter.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
Biological Effects of Radiation
Determination of Crystal Structures
Dose Response Curve: Conventional Versus Nonmonotonic
Absorption of Radiation