Related Experiment Video
Updated: May 30, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Improving initial acetabular component stability in revision total hip arthroplasty calcium phosphate cement vs
David J Jacofsky1, John D McCamley, Andrew M Jaczynski
1The Center for Orthopedic Research and Education (CORE) Institute.
Abstract:
A reproducible retroacetabular defect was created bilaterally in 9 cadaver pelves. The defects were filled with either an injectable, bioresorbable, calcium phosphate cement, or reverse-reamed cancellous allograft. An uncemented acetabular shell was impacted, followed by the placement of an appropriate liner. The pelves were then sectioned, and each half was loaded in a material testing machine to simulate walking on the construct over a several week period. The cement-filled defects lasted a greater number of cycles before failure and had greater cup stability and stiffness. The use of resorbable bone void filler for retroacetabular defects shows promise in this biomechanical analysis. Long-term clinical follow-up is warranted to track osseointegration of the implant and restoration of bone stock between this and other clinically accepted surgical techniques.