A Standardized Method for Measurement of Elbow Kinesthesia
Published on: October 10, 2020
Validity of goniometric elbow measurements: comparative study with a radiographic method
Julien Chapleau1, Fanny Canet, Yvan Petit
1Université de Montreal, Montreal, QC, Canada.
Clinical Orthopaedics and Related Research
|July 23, 2011
Summary
The universal goniometer is a simple tool for measuring elbow range of motion (ROM), but radiographic measurements offer higher precision. Goniometric measurements have a maximal error of about 10 degrees.
Area of Science:
- Orthopedics
- Biomechanical measurement
- Radiology
Background:
- The universal goniometer is a standard tool for assessing elbow range of motion (ROM) and carrying angle.
- Concerns exist regarding the intertester reliability of goniometric measurements.
Purpose of the Study:
- To evaluate the validity of goniometric elbow ROM measurements against radiographic assessments.
- To determine the reliability of both goniometric and radiographic measurement techniques.
Main Methods:
- Elbow ROM and carrying angle were measured in 51 healthy subjects (102 elbows) using a universal goniometer and radiographs.
- Goniometric measurements were taken three times by one observer; radiographic measurements were performed by two independent examiners.
- Paired t-tests, Pearson's correlation, and the Bland and Altman method were used for analysis.
Main Results:
- Intraclass correlation coefficients (ICC) indicated high reliability for both methods (goniometric: 0.945-0.973; radiographic: 0.980-0.991).
- Both methods showed correlation for total ROM in flexion and extension.
- Maximal errors for goniometric measurements were approximately 10.3° for extension, 7.0° for flexion, and 6.5° for carrying angle.
Conclusions:
- Goniometric and radiographic methods for elbow ROM assessment show differences but are correlated.
- The universal goniometer is a practical clinical tool, with a maximal error of ~10° for elbow ROM.
- Radiographic measurements are recommended for research requiring higher precision.