Implementation of multi-echo-based correlated spectroscopic imaging and pilot findings in human brain and calf muscle
Gaurav Verma1, Scott Lipnick, Saadallah Ramadan
1Department of Bioengineering, University of California, Los Angeles, California 90095-1721, USA.
Purpose:
To implement a spatially encoded correlated spectroscopic imaging (COSI) sequence on 3 Tesla (T) MRI/MR spectroscopy scanners incorporating four echoes to collect four phase-encoded acquisitions per repetition time (TR), and to evaluate the performance and reliability of this four-dimensional (4D) multi-echo COSI (ME-COSI) sequence in brain and calf muscle.
Materials And Methods:
Typical scan parameters for the 4D datasets were as follows: repetition time = 1500 ms, 2000 Hz bandwidth, 8 × 8 spatial encoding, one average, 64 Δt(1) increments and the scan duration was 25 min. The performance and test-retest reliability of ME-COSI were evaluated with phantoms and in the occipitoparietal brain tissues and calf of six healthy volunteers (mean age = 32 years old).
Results:
Regional differences in concentrations of lipids, creatine (Cr), choline (Ch), and carnosine (Car) were observed between spectra from voxels located in tibial marrow, tibialis anterior, and soleus muscle. Diagonal and cross-peak resonances were identified from several brain metabolites including N-acetyl aspartate (NAA), Ch, Cr, lactate (Lac), aspartate (Asp), glutathione (GSH), and glutamine\glutamate (Glx). Coefficients of variation (CV) in metabolite ratios across repeated measurements were <15% for diagonal and <25% for cross-peaks observed in vivo.
Conclusion:
The ME-COSI sequence reliably acquired spatially resolved 2D Correlated Spectroscopy (COSY) spectra demonstrating the feasibility of differentiating spatial variation of metabolites in different tissues. Multi-echo acquisition shortens scan duration to clinically feasible times.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies II: Ultrasonography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...