Prostate image-guided radiotherapy by megavolt cone-beam CT
Sergio Zucca1, Barbara Carau, Ignazio Solla
1Department of Radiooncology, Regional Oncological Hospital, Cagliari, Italy.
Summary
Megavolt cone-beam CT (MV-CBCT) effectively assesses soft tissue position for prostate radiotherapy setup verification. Soft tissue registration (STR) proved more accurate than bone markers for evaluating patient setup errors.
Area of Science:
- Medical Physics
- Radiation Oncology
- Imaging Technology
Background:
- Accurate patient setup is crucial for effective prostate radiotherapy.
- Traditional methods like electronic portal imaging (EPI) have limitations in verifying soft tissue positioning.
- Megavolt cone-beam CT (MV-CBCT) offers potential for improved setup verification.
Purpose of the Study:
- To evaluate the efficacy of MV-CBCT in assessing setup errors in prostate radiotherapy.
- To compare different MV-CBCT image registration methods (bone, soft tissue, automatic) against electronic portal imaging (EPI).
Main Methods:
- Weekly setup verification using EPI and MV-CBCT in 9 patients.
- MV-CBCT registration methods: manual bone (BMR), manual soft tissue (STR), and automatic (AR).
- Evaluation of shifts along anteroposterior (AP), craniocaudal (CC), and laterolateral (LL) axes.
- Analysis of calcification mismatch in 4 additional patients to assess registration accuracy.
Main Results:
- No significant differences between EPI and BMR in any axis within a 2 mm tolerance.
- Significant differences (p < 0.001) along the AP axis comparing EPI vs AR and EPI vs STR.
- Lower calcification mismatch (p < 0.05) along AP using STR compared to BMR or AR.
Conclusions:
- Bone markers are not reliable surrogates for prostate target position.
- MV-CBCT-based soft tissue registration (STR) is superior for verifying and correcting patient setup in prostate radiotherapy.
- MV-CBCT facilitates accurate assessment of soft tissue position for improved treatment accuracy.