Blood flow measurements during hemodialysis vascular access interventions--catheter-based thermodilution or Doppler
Søren T Heerwagen1, Marc A Hansen, Torben V Schroeder
1Department of Interventional Radiology, Rigshospitalet, Copenhagen, Denmark. soeren.heerwagen@rh.regionh.dk
Purpose:
To test the clinical performance of catheter-based thermodilution and Doppler ultrasound of the feeding brachial artery for blood flow measurements during hemodialysis vascular access interventions.
Methods:
Thirty patients with arteriovenous fistulas who underwent 46 interventions had access blood flow measured before and after every procedure. Two methods, catheter-based thermodilution and Doppler ultrasound, were compared to the reference method of ultrasound dilution. Catheter-based thermodilution and Doppler ultrasound were performed during the endovascular procedures while flow by ultrasound dilution was determined within three days of the procedure. The methods were compared using regression analysis and tested for systematic bias.
Results:
Failure to position the thermodilutional catheter correctly was observed in 8 out of 46 (17%) pre-intervention measurements. Post-intervention measurements and ultrasound measurements were feasible in all patients. The average level of agreement was good when comparing catheter-based thermodilution to ultrasound dilution. However, blood flow by ultrasound dilution may differ by ±130 mL/min (±22%) at a flow level of 600 mL/min by thermodilution. Results from Doppler ultrasound displayed a moderate level of agreement on average when compared to ultrasound dilution. Blood flow by ultrasound dilution may differ by ±160 mL/min (±27%) at a flow level of 600 mL/min by Doppler ultrasound. No systematic bias was detected by either method.
Conclusions:
On average, results from catheter-based thermodilution were more in agreement with results from the ultrasound dilution technique compared to Doppler ultrasound. However, considering the cost and the high technical failure rate of the thermodilutional system, we recommend the use of ultrasound.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Hemodialysis I: Introduction
Hemodialysis III: Nursing Management
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Hemodialysis II: Procedure and Complications
Imaging Studies VII: Vascular Imaging