Related Experiment Video
Updated: May 30, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Unusual false-positive case of urinary screening for buprenorphine
Giuseppe Lippi1, Araelsis Romero, Gianfranco Cervellin
1U.O. Diagnostica Ematochimica, Azienda Ospedaliero-Universitaria di Parma, Parma, Italy. glippi@ao.pr.it
Abstract:
Buprenorphine is a centrally acting analgesic drug that is administered for the management of opioid dependence and as an analgesic drug for the treatment of chronic pain. The growing use of this substance has determined an increased need for laboratory testing for either detection and confirmation of the illicit use or monitoring compliance as a substitution therapy for opioid dependence. We describe here the case of urinary sample adulteration with exogenous buprenorphine (6,952 ng/ml), which has led to afalse-positive immunoassay test result (14.9 ng/ml) on a subsequent sample due to a phenomenon of instrumental carry-over. This unusual case confirms the importance to take into account adulteration when screening urines for buprenorphine in patients undergoing substitution therapy for opioid dependence, routinely perform a confirmation assay on positive samples, and rule out instrumental carry-over.