Related Experiment Video
Updated: May 30, 2026

Measurement of Factor V Activity in Human Plasma Using a Microplate Coagulation Assay
Published on: September 9, 2012
[Experimental quality assessment of APTT-control reagent kits in the study of plasma with varying Factor VIII
Abstract:
An investigation into activated partial thromboplastin time (APTT) in the plasma samples with varying Factor VIII activities has determined the following parameters of the quality of APTT-control reagent kits: coefficient of variation, linearity, sensitivity, and detection. The kit is shown to meet GOST P 51352-99 requirements. The determination of the validity coefficient of approximation has proven a directly proportional functional relationship of the APTT values to the Factor VIII activity in the bilogarithmic coordinate system. The kit may be used to diagnose hemophilia A and Willebrand disease.