L P J Teunissen1, A de Haan, J J de Koning
1TNO, Human Performance, PO Box 23, 3769 ZG, Soesterberg, The Netherlands.
Abstract:
Aural canal temperature measurement using an ear mould integrated sensor (T(ac)) might be a method suited for continuous non-invasive core temperature estimation in operational settings. We studied the effect of ambient temperature, wind and high intensity exercise on T(ac) and its ability to predict esophageal (T(es)) and rectal temperatures (T(re)). Seven subjects performed a protocol of rest at 21, 10 and 30 °C, followed by exercise and recovery at 30 °C. The subjects performed the protocol twice: with and without face-wind from halfway through the 30 °C rest period. Extra auricle insulation was applied at one side. Ambient temperature changes affected T(ac) significantly, while T(es) and T(re) remained stable. Insulating the auricle reduced but did not abolish this effect. Wind had an immediate cooling effect on T(ac) independent of auricle insulation. During exercise and recovery in 30 °C, T(ac) provided acceptable group predictions of T(re) in trials without wind (bias: -0.66 ± 0.21 °C covered, -1.20 ± 0.15 °C uncovered). Bias was considerably higher with wind, but variability was similar (-1.73 ± 0.11 °C covered, -2.49 ± 0.04 °C uncovered). Individual predictions of T(es) and T(re) showed more variation, especially with wind. We conclude that T(ac) may be used for core temperature assessment of groups in warm and stable conditions.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...