Injected contrast study fails to demonstrate catheter-pump connector tear
Linda E Krach1, Michael D Partington
1Department of Pediatric Rehabilitation Medicine, Gillette Children's Specialty Healthcare, St. Paul, MN, USA Department of Physical Medicine and Rehabilitation, University of Minnesota Medical School, Minneapolis, MN, USA.
A torn intrathecal baclofen (ITB) pump catheter connector caused a pediatric patient's increased tone. Diagnosis was confirmed via imaging and clinical response, leading to successful surgical repair and improved ITB therapy.
Area of Science:
- Pediatric Neurology
- Neurosurgery
- Rehabilitation Medicine
Background:
- Intrathecal baclofen (ITB) therapy is crucial for managing severe spasticity in conditions like cerebral palsy (CP).
- Patients on ITB therapy may experience increased tone, necessitating investigation into potential complications.
- Device-related issues, such as catheter or pump malfunction, are common causes of treatment failure.
Purpose of the Study:
- To present a case of a pediatric patient with CP experiencing increased tone due to a malfunctioning ITB system.
- To highlight the diagnostic process and successful management of a rare ITB pump-catheter connector tear.
Main Methods:
- A 11-year-old boy with CP, shunted hydrocephalus, and ITB therapy for hypertonicity presented with worsening tone.
- Diagnostic evaluation included clinical assessment, review of imaging (X-rays), and response to a test bolus of ITB.
- Surgical exploration was performed to identify and repair the suspected device failure.
Main Results:
- X-ray findings suggested a tear in the pump-catheter connector sleeve.
- A test bolus of ITB provided temporary symptom relief.
- Surgery confirmed a torn pump-catheter connector, which was replaced, resolving the patient's symptoms and allowing for optimized ITB dosage.
Conclusions:
- Increased ITB dosage requirements in patients with CP often indicate a catheter or pump-related problem.
- Palpable fluid in the pump pocket is a significant indicator of a potential catheter/pump connector tear.
- Accurate diagnosis can be achieved through careful clinical evaluation, radiographic review, and assessment of treatment response.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Cardiac Catheterization I: Pre-Procedure Overview