A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
True bipolar and integrated bipolar sensing and detection by implantable defibrillators
Trevor A Verga1, Jeffrey M Gillberg, Richard M Greenberg
1Department of Cardiac Electrophysiology, Temple University Hospital, Philadelphia, Pennsylvania 19140, USA.
This study found no significant difference in R-wave amplitude or ventricular fibrillation detection time between true bipolar and integrated bipolar configurations in implantable defibrillators. These findings suggest similar performance for both sensing configurations in Medtronic devices.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Implantable defibrillators utilize either true bipolar or integrated bipolar configurations for sensing and detection.
- Newer Medtronic generators offer adjustable configurations for sensing functions.
Purpose of the Study:
- To compare the sinus rhythm R-wave amplitude and ventricular fibrillation (VF) detection time between true bipolar and integrated bipolar configurations.
- To evaluate acute and chronic phase performance of these sensing configurations in implantable defibrillators.
Main Methods:
- Twenty-eight patients were assessed in the acute phase, with 15 evaluated in the chronic phase.
- R-wave amplitude and VF detection time were measured for both true bipolar and integrated bipolar configurations using specific Medtronic generators and leads.
- Electrograms were recorded, and detection time was measured from T shock to VF onset.
Main Results:
- No significant difference was observed in acute-phase R-wave amplitude (13.9 ± 7.1 mV vs. 13.6 ± 6.9 mV, p = 0.38) between the two configurations.
- Ventricular fibrillation detection times were also comparable between true bipolar (3.12 ± 0.39 seconds) and integrated bipolar (3.17 ± 0.39 seconds) configurations (p = 0.52).
Conclusions:
- Sensing and detection performance were not significantly different between true bipolar and integrated bipolar configurations.
- The findings apply to the specific leads and generators tested, indicating comparable efficacy for both configurations.