The ArcCHECK diode array for dosimetric verification of HybridArc
A L Petoukhova1, J van Egmond, M G C Eenink
1Radiotherapy Centre West, PO Box 432, NL-2501 CK, The Hague, The Netherlands. a.petoukhova@mchaaglanden.nl
This study validates HybridArc treatment planning accuracy using Monte Carlo (MC) calculations, showing good agreement with measurements. Pencil Beam (PB) calculations were less accurate, highlighting MC
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Planning
Background:
- HybridArc is an advanced radiotherapy technique combining dynamic conformal arcs with inverse planning.
- Accurate dose calculation is crucial for effective and safe radiation delivery in complex treatment modalities.
Purpose of the Study:
- To assess the dosimetric accuracy of the HybridArc treatment planning system (iPlan RT Dose 4.5).
- To compare the performance of Monte Carlo (MC) and Pencil Beam (PB) dose algorithms for HybridArc plans.
- To evaluate HybridArc plans for both intracranial and extracranial sites.
Main Methods:
- HybridArc plans were created for various anatomical sites.
- Plans were recalculated using MC and PB algorithms in iPlan RT Dose.
- Dosimetric measurements were performed using a four-dimensional diode array (ArcCHECK) and an ionization chamber with a Novalis accelerator (6 MV).
- Calculated and measured dose distributions were compared using 2% dose difference and 2 mm distance-to-agreement criteria.
Main Results:
- Monte Carlo (MC) algorithm calculations showed good agreement with ArcCHECK measurements for HybridArc plans.
- Pencil Beam (PB) algorithm calculations demonstrated significant differences compared to ArcCHECK measurements.
- The MC algorithm was found to be superior to the PB algorithm for HybridArc plan calculations.
- A large arc step size (10°) in PB calculations was identified as a potential limitation.
Conclusions:
- The Monte Carlo (MC) algorithm provides accurate dose calculations for the HybridArc treatment modality.
- The Pencil Beam (PB) algorithm is less suitable for HybridArc dose calculations, necessitating improvements in angular resolution.
- HybridArc shows dosimetric accuracy suitable for clinical implementation when planned and calculated with appropriate algorithms.