Related Experiment Video
Updated: May 30, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
Published on: October 23, 2018
Development of a compact fast CCD camera and resonant soft x-ray scattering endstation for time-resolved pump-probe
D Doering1, Y-D Chuang, N Andresen
1Engineering Division, Lawrence Berkeley National Laboratory, 1 Cyclotron Rd., Berkeley, California 94720, USA. ddoering@lbl.gov
Abstract:
The designs of a compact, fast CCD (cFCCD) camera, together with a resonant soft x-ray scattering endstation, are presented. The cFCCD camera consists of a highly parallel, custom, thick, high-resistivity CCD, readout by a custom 16-channel application specific integrated circuit to reach the maximum readout rate of 200 frames per second. The camera is mounted on a virtual-axis flip stage inside the RSXS chamber. When this flip stage is coupled to a differentially pumped rotary seal, the detector assembly can rotate about 100°/360° in the vertical/horizontal scattering planes. With a six-degrees-of-freedom cryogenic sample goniometer, this endstation has the capability to detect the superlattice reflections from the electronic orderings showing up in the lower hemisphere. The complete system has been tested at the Advanced Light Source, Lawrence Berkeley National Laboratory, and has been used in multiple experiments at the Linac Coherent Light Source, SLAC National Accelerator Laboratory.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
10:18Dynamic Pore-scale Reservoir-condition Imaging of Reaction in Carbonates Using Synchrotron Fast Tomography
Published on: February 21, 2017